Human Cytokine Array

Sample Testing Service

Gentel's Sample Testing Service for human cytokines offers a multiplexed sandwich immunoassay for the simultaneous quantitative measurement of 17 analytes. Our quantitative microspot immunoassays are developed and validated in serum and plasma using acceptance criteria established by the FDA for bioanalytical methods. Simply select your proteins of interest, ship samples to Gentel, and receive a confidential data report within 14 business days.

With the Sample Testing Service for Cytokines, you obtain:

  • Reliable, quantitative results for 17 analytes per sample
  • Exceptional sensitivity: low- to sub-pg/ml levels
  • Excellent reproducibility: ≤15%CVs
  • High selectivity: ≤5% cross reactivity
  • Accurate results: dilutional recovery of 85-115%
  • Low sample volume (100μl) consumption
ANALYTE DETECTION LIMITS (pg/mL) 3-DAY VARIABILITY ‡ % DILUTION RECOVERY : SERUM MATRIX ♦
LOD LOQ AVERAGE %CV± SD AVERAGE r2 ± SD SPIKED PROTEIN (pg/mL) % DILUTIONAL RECOVERY
GM-CSF 0.13 1.93 6.7 ± 1.8 0.998 ± 0.001 150 104
IFNγ 0.15 1.87  4.7 ± 1.8 0.999 ± 0.000 67 98
IL-1β 0.09 0.71  4.2 ± 0.4 1.000 ± 0.001 40 100
IL-2 0.59 6.81  9.7 ± 1.4 0.998 ± 0.001 50 105
IL-3 0.90 6.96  6.4 ± 2.4 0.998 ± 0.001 200 96
IL-4 0.20 1.72  6.5 ± 2.4 0.999 ± 0.000 33 95
IL-5 0.64 5.70  6.5 ± 1.3 0.997 ± 0.002 20 100
IL-6 0.02 0.37  9.1 ± 1.7 0.998 ± 0.001 20 106
IL-7 0.05 1.07 15.3 ± 7.6 0.997 ± 0.002 20 94
IL-8 0.03 0.73 14.6 ± 4.0 0.997 ± 0.002 27 108
IL-10 0.00 0.09  7.9 ± 2.4 0.999 ± 0.001 40 98
IL-12 2.08 42.35  9.8 ± 2.6 0.998 ± 0.001 400 99
IL-13 0.05 0.76  7.7 ± 2.9 0.999 ± 0.001 160 100
MCP-1 1.20 6.58 11.5 ± 2.0 0.998 ± 0.001 40 105
TNFα 0.23 2.57 14.4 ± 4.8 0.996 ± 0.003 17 102
TNFβ 0.12 1.53  6.7 ± 3.1 0.999 ± 0.001 67 102
VEGF 0.15 1.49  9.6 ± 1.9 0.999 ± 0.001 33 96

†‡ Two standard curves were prepared each day for three consecutive days (n = 6). For each individual analyte the %CV at each concentration was averaged. The average LOD, LOQ and %CVs for each analyte is shown.

† LOD = Limit of Detection meaning the lowest level of analyte that can be detected. LOQ = Limit of Quantitation meaning the lowest level of anlalyte that can be reliably measured with acceptable accuracy and precision.

‡%CV = Percent correlation of variation. r2 = Pearson’s correlation coefficient.

♦Human serum was diluted 1:4 with fetal calf serum and spiked with an analytical cocktail of recombinant proteins. Each protein was measured in triplicate each day for three consecutive days (n = 9). The differences between the percentages of the measured concentration and the calculated concentrations are shown.